This standard specifies the safety technical requirements for coal industry analyzers in measuring the moisture, ash and volatile matter of coal and coke, and is applicable to laboratory environments with a rated voltage not exceeding 600V.
Key indicators of electrical safety
| Protection type | Electrical clearance (mm) | Creep distance (mm) | Test voltage (AC) |
|---|---|---|---|
| Basic insulation | ≥0.2 | ≥1.0 | 1500V |
| Double insulation | ≥0.4 | ≥2.0 | 3000V |
Note: The minimum electrical clearance under pollution degree 2 is 0.2mm, and high-voltage components must pass the 1.2/50μs pulse voltage test.
Mechanical safety design requirements
Protection of moving parts must meet the following requirements:
- Rotating parts should be equipped with protective covers with an opening size of ≤12mm
- Equipment with a mass of ≥18kg must be equipped with a special handling device
- Static pressure bearing capacity of the shell is 30N, and dynamic impact energy is 5J
Fire prevention and temperature control
Key fire prevention measures:
- Non-metallic shell materials must meet the flame retardant grade of GB/T 11020 FV-1
- Microwave leakage ≤50W/m² (50mm from the surface)
- Touchable surface temperature ≤70℃ (metal)/80℃ (non-metal)
Implementation suggestions
Purchase and acceptance
It is recommended to give priority to products that have passed CNAS certification. During acceptance, the following checks must be checked:
- Protective grounding impedance ≤ 0.1Ω
- Double insulation mark (return symbol)
- Verification of the effectiveness of the safety interlock device
Daily maintenance
Every quarter, the following should be carried out:
- Electrical safety test (insulation resistance ≥ 10MΩ)
- Lubrication check of mechanical parts
- Microwave sealing strip integrity test