Sign In |Help & Support
ALL SECTORS
  • ALL SECTORS
  • GB(National Standard)
  • CB(Shipping)
  • CECS(Engineering Construction)
  • CJ(Urban Construction)
  • CY(News and Publication)
  • DB(Provincial Standard)
  • DL(Electricity & Power)
  • DZ(Geology & Mineralogy)
  • FZ(Spinning & Textile)
  • GA(Public Security)
  • HB(Aviation)
  • HG(Chemical Industry)
  • HJ(Environmental Protection)
  • JB(Machinery)
  • JC(Building Materials)
  • JG(Building & Construction)
  • JJ(Metering)
  • JT(Highway & Transportation)
  • LY(Forestry)
  • MT(Coal)
  • NB(Energy)
  • NY(Agriculture)
  • QB(Light Industry)
  • QC(Automobile & Vehicle)
  • QJ(Aerospace)
  • SH(Petrochemical)
  • SJ(Electronics)
  • SL(Water Resources)
  • SN(Commodity Inspection)
  • SY(Oil & Gas)
  • TB(Railway & Train)
  • YB(Ferrous Metallurgy)
  • YC(Tobacco)
  • YD(Telecommunication)
  • YY(Medical Device)
Database: 365,228(8 Aug 2026)
optical system design standard requirements standard technical indicatorsgb/t standard pyrheliometer introduction analysis standard requirement requirementsinternational standard iso address data type jc4520 cmos self-generating electricity rate
GB/T 33706-2017 in English

GB/T 33706-2017 in English

VALID

Standard pyrheliometer

  • Issued on:2017-05-12
  • Implemented on:2017-12-01
  • File Format:PDF
  • Delivery:Via email within 5 business days
Price(USD): $370.00
$359.00

《GB/T 33706-2017标准直接辐射表》由TC507(全国气象仪器与观测方法标准化技术委员会)归口,主管部门为中国气象局。


Introduction

Analysis of the core framework of the standard

Technical indicatorsGB/T 33706 requirementsInternational standard ISO 9060Differences in test methods
Response time≤10s≤15sDynamic step test method
Zero point offset≤2W/m²≤5W/m²Temperature cycle test
Spectral range300-3000nm280-3000nmSpectrophotometer method

Detailed explanation of key technical requirements

Optical system design

Standard requirements: The geometric parameters of the collimation tube must meet the following requirements:

  • Half-opening angle θ: 2.5°±0.1°
  • Bevel angle δ: 1°±0.1°

This design ensures accurate measurement of the 5.7° field of view and effectively eliminates stray light interference.


Environmental adaptability requirements

According to Appendix A, the instrument must pass the following rigorous tests:

  1. Temperature cycle: -10℃→40℃ (5K/h change rate)
  2. Alternating damp heat: 12h cycle (40℃/95%RH)
  3. Low pressure test: 450hPa for 4 hours

Case: The actual measurement of a certain model of direct radiation meter on the Qinghai-Tibet Plateau (3650m above sea level) shows that the temperature response error is ≤0.8%, which is better than the standard requirement of ±1%.


Implementation suggestions

Calibration cycle management

The standard stipulates that the calibration cycle of direct radiation meter is 2 years, but it is recommended to shorten the cycle in the following situations:

  • Highly polluted environment (PM2.5>75μg/m³)
  • Coastal high salt fog area
  • In strong radiation area with annual radiation>1800kWh/m²

Maintenance points

Key maintenance steps:

  1. Check the cleanliness of the quartz window daily (wipe with anhydrous ethanol)
  2. Verify the desiccant status every month (replace when the color-changing silica gel>1/3 turns red)
  3. Check the coating integrity of the sensing surface every quarter (blackness coefficient change ≤3%)