GB/T 38499-2020 in English
VALIDEvaluation method for stability of disinfectant
- Issued on:2020-03-06
- Implemented on:2020-10-01
- File Format:PDF
- Delivery:Via email within 1~3 business days
$156.00
《GB/T 38499-2020消毒剂稳定性评价方法》由628(国家疾病预防控制局)归口,主管部门为国家疾病预防控制局。
Introduction
Analysis of the core content of the standard
Comparison of test methods
| Test type | Temperature conditions | Humidity requirements | Light intensity | Maximum prediction period |
|---|---|---|---|---|
| Accelerated test | 54℃±2℃ | Liquid not required Solid 75%±5% | - | 14 days→12 months |
| Long-term test | 25/30℃±2℃ | 60±10%/65±5% | - | 12-24 months |
| Strong light irradiation | 25℃±2℃ | 60±10% | 4500lx±500lx | 10 days |
Key technical points
1. Sample requirements
Must use shaped packaging disinfectant, 3 independent production batches, and large barrel packaging must use simulated small packaging with the same material.
2. Instrument accuracy
- Constant temperature and humidity chamber: temperature ±2℃, humidity ±5%
- Lighting test chamber: illumination control 4500lx±500lx
3. Evaluation criteria
- Unstable ingredients (peracetic acid, etc.): content decrease ≤15%
- Other ingredients: content decrease ≤10%
- Relative average deviation between batches ≤5%
Implementation suggestions
Enterprise execution path
- New raw materials must be tested: New active ingredients must be tested for strong light exposure
- Double verification: Accelerated test results must be confirmed by long-term testing
- Packaging labeling: 25°C storage conditions must be marked with the "Avoid high temperature" warning
Recommended testing frequency
| Test phase | Test time point |
|---|---|
| Long-term testing | 0, 3, 6, 9, 12 months (extended to 24 months) |
| Accelerated testing | Immediate testing after the test |
Standard evolution analysis
This standard systematically integrates the three major method systems for disinfectant stability evaluation for the first time. Compared with the 2002 version of the Technical Specification for Disinfection:
- Newly added strong light irradiation test as a mandatory test for photostability
- Refine the correspondence between accelerated test and shelf life
- Clarify the priority principle between chemical method and microbiological method

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

GB 38456-2020 in English
Hygienic requirements for antibacterial and bacteriostatic lotion
2020-11-17 -

GB/T 26367-2020 in English
Hygienic requirements for biguanides disinfectants
2020-06-02 -

GB/T 36758-2018 in English
Hygienic requirements for disinfectants with chlorine
2018-09-17 -

GB/T 27947-2020 in English
Hygienic requirements for phenol disinfectant
2020-06-02 -

GB/T 38496-2020 in English
Toxicological procedures and methods of safety evaluation for disinfectant
2020-03-06 -

GB/T 33417-2016 in English
Test method of biological indicator for hydrogen peroxide vapour sterilization processes
2016-12-30