GB/T 44423-2024 in English
VALIDGeneral requirements for near-infrared brain function rehabilitation assessment equipment
- Issued on:2024-08-23
- Implemented on:2024-12-01
- File Format:PDF
- Delivery:Via email within 5 business days
$359.00
《GB/T 44423-2024近红外脑功能康复评估设备通用要求》由TC148(全国残疾人康复和专用设备标准化技术委员会)归口,主管部门为民政部。
Introduction
Technical requirements and applications of near-infrared brain function rehabilitation assessment equipment
| Technical indicators | Requirement description | Application scenarios |
|---|---|---|
| Emission wavelength | The equipment should emit no less than two wavelengths, one of which has a peak wavelength of 690 nm~785 nm and the other has a peak wavelength of 825 nm~900 nm. | Ensure that the equipment can accurately measure the changes in hemoglobin concentration in brain tissue and is suitable for the assessment of brain regions at different depths. |
| Temporal resolution | When measuring all channels, the maximum temporal resolution should be no less than 10 Hz. | Support high-frequency data acquisition, suitable for real-time monitoring of dynamic changes in brain function. |
| Receiver probe dynamic range | The device should provide a receiver probe dynamic range of no less than 100 dB. | Ensure that the device can work effectively in both high-intensity and low-intensity environments, and is suitable for brain function assessment in complex environments. |
Standard formulation background and technology evolution analysis
The standard formulation of near-infrared brain function rehabilitation assessment equipment is based on the following background:
- Technological progress:With the rapid development of near-infrared spectroscopy technology, its application in brain function assessment is becoming increasingly widespread.
- Clinical needs:Patients with brain injury, developmental disorders, and motor, cognitive, and mental disorders have an urgent need for accurate rehabilitation assessment.
- Industry specifications:To ensure equipment performance and safety, regulate the market and promote the healthy development of the industry.
Implementation Suggestions
To ensure the effective application of near-infrared brain function rehabilitation assessment equipment, the following implementation suggestions are proposed:
- Training and operation:Systematic training for medical staff to ensure the correct use of the equipment and its software.
- Quality control:Regularly calibrate the equipment to ensure that the technical parameters meet the standard requirements.
- Data management:Establish a complete data management system to ensure patient privacy and data security.

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

GB/T 24434-2025 in English
Commode chairs
2025-06-30 -

GB/Z 44426.7-2026 in English
Assistive products for tissue integrity when lying down—Part 7:Foam properties, characteristics and performance
2026-01-28 -

GB/T 45265-2025 in English
General technical requirements for additive manufacturing of lower limb prostheses
2025-02-28