JB/T 12934-2016 in English
VALIDFormaldehyde gas sensor
- Issued on:2016-10-22
- Implemented on:2017-04-01
- File Format:PDF
- Delivery:Via email within 1~3 business days
Price(USD):
$329.00
$320.00
$320.00
| Standard No: | JB/T 12934-2016 |
| Document status: | VALID |
| Title in English: | Formaldehyde gas sensor |
| Title in Chinese: | 甲醛气体传感器 |
| Language: | English |
| File Format: | Electronic (PDF) |
| Delivery: | Via email within 1~3 business days |
| Issued on: | 2016-10-22 |
| Implemented on: | 2017-04-01 |
| ICS Classification: | 13.040.20-Ambient atmospheres |
| Chinese Classification: | N05-Material and component for instrument and meter |
| Professional Classification: | JB-Machinery |
Introduction
Analysis of the core content of the standard
| Technical indicators | Electrochemical sensor | Semiconductor sensor |
|---|---|---|
| Response time | ≤120s | ≤120s |
| Basic error | ≤10% | ≤10% |
| Operating temperature | -20℃~50℃ | -20℃~50℃ |
Key technical requirements
The standard stipulates that formaldehyde gas sensors must meet 20 performance indicators in five categories:
- Environmental adaptability: Pass rigorous tests such as -25℃~55℃ temperature storage, alternating humidity and heat, vibration and shock
- Accuracy index: Long-term sensitivity drift ≤18%, zero drift ≤0.5×10-6
- Dynamic characteristics: Response/recovery time does not exceed 120 seconds
Test method system
The standard establishes a three-level verification system:
- Basic test: Sensitivity is determined using the step gas concentration change method
- Environmental test: 12 environmental stress tests are carried out in accordance with the GB/T 2423 series standards
- Long-term stability: Drift characteristics verified through 720 hours of continuous testing
Recommendations for standard implementation
Professional suggestions for standard implementation:
| Application scenarios | Implementation points |
|---|---|
| Industrial production environment | Give priority to electrochemical principle sensors, and conduct alternating humidity and heat verification regularly |
| Civilian testing equipment | Focus on monitoring long-term zero drift indicators, and it is recommended to calibrate every 6 months |
Sample only — not a preview of JB/T 12934-2016

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

JB/T 14855-2025 in English
Electrochemical alcohol gas sensor
2025-05-09