VALIDRehabilitation training devicesGeneral specifications of active upper limb training devices
- Issued on:2020-10-23
- Implemented on:2020-10-23
- File Format:PDF
- Delivery:Via email within 1~3 business days
Price(USD):
$250.00
$243.00
$243.00
| Standard No: | MZ/T 152-2020 |
| Document status: | VALID |
| Title in English: | Rehabilitation training devicesGeneral specifications of active upper limb training devices |
| Title in Chinese: | 康复训练器械 主动式上肢关节训练器通用技术条件 |
| Language: | English |
| File Format: | Electronic (PDF) |
| Delivery: | Via email within 1~3 business days |
| Issued on: | 2020-10-23 |
| Implemented on: | 2020-10-23 |
| ICS Classification: | 11.040.99-Other medical equipment |
| Chinese Classification: | C30-Medical apparatus and devices in general |
| Professional Classification: | MZ-Civil Affairs |
| Related Keywords: | active upper limb training devices introduction analysis
resistance training damping torque requirements resistance training combined training upper limb joint trainer |
| Related Topics: | recover
train active degausser mz initiative Rehabilitation Rehabilitation Upper Limb Rehabilitation Training Ministry of Justice Rehabilitation Training Information System Technology t/caqi 152-2020 |
本标准规定了主动式上肢关节训练器的分类、技术要求、试验方法、标志、使用说明书、包装、运 输和贮存。 本标准适用于对上肢关节作主动训练的康复器械。 本标准不适用于对上肢关节作主/被动混合训练的康复器械。
Introduction
Analysis of the core content of the standard
Active upper limb joint trainer is defined as a rehabilitation device that can provide damping force for resistance training by actively moving the user's limbs, mainly for single or combined training of the shoulder, elbow, wrist and finger joints.
Comparison of key technical requirements
| Technical dimensions | Self-standing requirements | Fixed requirements |
|---|---|---|
| Stability | Must pass the tipping stability test (GB 24436 6.3.2) | Installation firmness test (5.11.2) |
| Range of motion | Shoulder flexion ≥ 90°, wrist dorsiflexion ≥ 70° (Table 1) | |
| Safety clearance | The distance between moving parts is >60mm (except for special circumstances in 5.3.4) | |
Key parameters of resistance training
Damping torque requirements:
• Shoulder/elbow joint ≥15N·m
• Wrist joint ≥10N·m
• Finger joint ≥5N·m (Table 2)
• Shoulder/elbow joint ≥15N·m
• Wrist joint ≥10N·m
• Finger joint ≥5N·m (Table 2)
Rehabilitation training equipmentThe weight system must meet the following requirements:
• Finger training weight 0.1-0.4kg (error ±3%)
• Other joint weights 0.5-2.5kg (5.5.4)
• Finger training weight 0.1-0.4kg (error ±3%)
• Other joint weights 0.5-2.5kg (5.5.4)
Safety test specifications
- Fatigue test: Complete 100,000 cycles at a frequency of 30 times/minute (6.11)
- Static load test: The cable must withstand 6 times the rated resistance (5.9.1)
- Handle test: Meet the anti-slip requirements of GB 24436 (5.6.2)
Implementation suggestions
Clinical selection notes:
1. Fixed equipment needs to check the wall's load-bearing capacity (5.11.1)
2. Children's equipment needs to ensure that the finger gap is <4mm or >25mm (5.3.4f)
3. Resistance equipment should be equipped with a torque test report (6.6.3)
1. Fixed equipment needs to check the wall's load-bearing capacity (5.11.1)
2. Children's equipment needs to ensure that the finger gap is <4mm or >25mm (5.3.4f)
3. Resistance equipment should be equipped with a torque test report (6.6.3)
Sample only — not a preview of MZ/T 152-2020

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

MZ/T 157-2020 in English
Manual transfer trolley
2020-10-23 -

MZ/T 153-2020 in English
Rehabilitation training devices-Parallel bars
2020-10-23 -

MZ/T 154-2020 in English
Mammary prostheses
2020-10-23 -

MZ/T 159-2020 in English
Vertical installed combined support handrails
2020-10-23 -

MZ/T 158-2020 in English
Assist armrest fixed on toilet
2020-10-23 -

MZ/T 161-2020 in English
General technical requirements for powered air-purifying respirator
2020-10-23