VALIDDetermination of malachite green in water and sediment from the aquaculture environment- -High performance liquid chromatography
- Issued on:2019-08-01
- Implemented on:2019-11-01
- File Format:PDF
- Delivery:Via email within 1~3 business days
$156.00
Introduction
Analysis of the core content of the standard
This standard adopts a high performance liquid chromatograph in conjunction with a fluorescence detector to establish a detection method for malachite green in aquaculture environments (water bodies/sediment), with a quantitative limit of 0.1 μg/L (water bodies) and 1.0 μg/kg (sediment).
Comparison of technical methods
| Detection object | Pretreatment method | Chromatographic conditions | Limit of quantification |
|---|---|---|---|
| Aquaculture water | Potassium borohydride reduction + MCX column purification | C18 column, acetonitrile-ammonium acetate mobile phase | 0.1 μg/L |
| Aquaculture sludge | Ultrasonic extraction + alumina purification | Same water conditions | 1.0 μg/kg |
Key operating points
Sample processing specifications
Water samples must be filtered through a 0.45 μm filter membrane, and potassium borohydride solution must be added to reduce the latent malachite green; sediment samples must be extracted with a dichloromethane-acetonitrile mixture using an ultrasonic cleaner.
Chromatographic parameter setting
- Chromatographic column: C18 column (250mm×4.6mm, 5μm)
- Mobile phase: acetonitrile-0.125mol/L ammonium acetate buffer (80:20)
- Detection wavelength: 265nm (excitation)/360nm (emission)
Standard implementation recommendations
1. The laboratory should regularly verify the recovery rate of the solid phase extraction device, and it is recommended to perform matrix spike experiments for each batch of samples;
2. Note that the standard solution must be prepared and used and stored away from light for no more than 3 months;
3. The water content of the sediment test must be measured simultaneously for result correction.
Technology Evolution Analysis
Compared with the early colorimetric method, this standard has the following advantages:
• Sensitivity is improved by more than 100 times
• Malachite green and its metabolites can be detected simultaneously
• MCX column purification technology is introduced to reduce matrix interference