WS/T 224-2018 in English
VALIDPerformance verification of vacuum tubes for venous blood specimen
- Issued on:2018-04-27
- Implemented on:2018-11-01
- File Format:PDF
- Delivery:Via email within 1~3 business days
$311.00
本标准规定了真空采血管性能要求及验证方法。
本标准适用于临床实验室在正式使用新采血管前对其进行性能验证。对正在使用的采血管的检测结果有疑问、查找原因可能与采血管相关时,也可采用本标准对采血管进行性能验证。
Introduction
Core Changes of the Standard Revision
| Comparison Dimensions | 2002 Edition | 2018 Edition |
|---|---|---|
| Standard Name | Vacuum Blood Collection Tubes and Their Additives | Performance Verification of Vacuum Blood Collection Tubes |
| Technical Indicators | Mainly Manufacturer Testing | Four New Clinical Evaluation Indicators |
| Centrifugal Force Requirements | Value Not Cleared | Increased to 3000g |
Key Performance Verification Points
1. Pumping Volume Verification
Use the gravimetric analysis method, and the relative bias is required to be within ±10%. Verification steps:
- Use a 20G standard blood collection needle to collect water
- Weigh 30 samples and calculate the average value
- Determine according to the formula RD=(DV-NLC)/NLC×100%
2. Tube Strength Test
The new standard increases the centrifugal force requirement from unclear in the original standard to 3000g centrifugation for 10 minutes. Please note the following during verification:
- Sample volume: 30 tubes filled with water to the scale line
- A centrifugal radius of 17cm corresponds to 4000r/min
- Observe tube leakage
Clinical Evaluation Innovation Indicators
Serum Separation Tube Verification Case
Testing on 40 samples revealed the following:
- Batch Management: New batches of blood collection tubes require full verification
- Joint Verification: 4.4-4.6 indicators can be tested simultaneously
- Data Retention: Original weighing records should be kept for at least 2 years
- Whether the number of shakes after blood collection meets the requirements
- Whether the standing time reaches 30 minutes
- Whether the centrifugation parameters are accurate
Implementation Suggestions
Verification Process Optimization
Troubleshooting
When fibrin adhesion occurs, check:

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

WS/T 491-2024 in English
Guidelines for non-specific syphilis antibody testing
2024-05-09 -

WS/T 514-2017 in English
Establishment and verification of detection capability for clinical laboratory measurement procedures
2017-01-15 -

WS/T 402-2012 in English
Define and determine the reference intervals in clinical laboratory
2012-12-24