YY 0451-2023 in English
TO BE VALIDPortable infusion devices for single use--Non electrically driven
- Issued on:2023-09-05
- Implemented on:2026-09-15
- File Format:PDF
- Delivery:Via email within 1~3 business days
$243.00
本文件规定了一次性使用便携式输注泵-非电驱动(以下简称输注泵)的基本要求和相应的试验方法。
本文件适用于神经、血管内或皮下应用的可持续(固定的或可调节)给液和/或自控给液的输注泵。
本文件不适用于:
GB 9706.224所涵盖的电动或电控输注泵;
供单个患者使用的,用于输送YY/T 1768系列所涵盖的离散(丸剂)药液的装置;
植入器械;
肠给养泵;
经皮给液装置;
输液动力不是由装置本身提供或由患者主动干预获得动力的装置(例如,仅由重力作为动力的装置)。
Introduction
Background of Standard Revision and Technological Evolution
As the industry standard for non-electrically driven portable infusion pumps, YY 0451-2023 replaces the 2010 version and achieves three major upgrades: refined technical indicators (such as flow rate deviation tightened from ±20% to ±15%), comprehensive safety requirements (newly added particle pollution and chemical performance testing), and international integration (modification to adopt ISO 28620:2020). This standard is applicable to neural/vascular/subcutaneous continuous drug delivery scenarios, and explicitly excludes gravity infusion devices and implantable devices.
Comparison of core technical requirements
| Project | YY 0451-2010 | YY 0451-2023 | Change analysis |
|---|---|---|---|
| Flow accuracy | ±20% | ±15% (fixed flow) ±20% (adjustable flow) | Increased clinical drug delivery accuracy requirements |
| Filter pore size | Unspecified | ≤5μm (must comply with YY 0286.1) | Prevent the risk of particulate embolism |
| Chemical properties | Not required | New Chapter 4.5 (6 items including heavy metals/UV absorbance) | Strengthening of material safety control |
Analysis of key test methods
1. Flow measurement (Clause 6.2)
Adopt the dynamic weighing method, with the following requirements:
- Average flow deviation ≤±15% (adjustable ±20%)
- Instantaneous flow fluctuation within 80% of the nominal capacity period ≤±50%
The test must be carried out after completing the compression (150N/5s) and tensile (15N/15s) tests to simulate the actual usage conditions.
2. Validation of automatic liquid feeding system (clauses 6.7-6.8)
New models must pass triple validation:
- Standard trigger: liquid feeding volume ≤115% of nominal value
- Fast trigger (50% filling time): liquid feeding volume <75% of nominal value
- Full filling trigger: liquid feeding volume 50%-115% of nominal value
Implementation suggestions
Key points for corporate compliance
- Material selection: materials that have passed the biocompatibility evaluation of GB/T 16886.1 are preferred
- Connector design: must comply with the locking requirements of YY/T 0916.6/0916.7, and it is recommended to mark the connector type
- Process validation: focus on monitoring the sterilization process (EO residue ≤10μg/g) and particulate contamination control (as per Appendix B)
Clinical use tips
According to the requirements of Chapter 8 of the standard, manufacturers must clearly state in the instructions:
- List of contraindicated drugs (such as incompatible chemotherapy drugs)
- Temperature effect coefficient on flow rate (rate of change per °C)
- Intravenous infusion warning label for neurological instruments
Typical case: Development of prefilled infusion pumps
A company's prefilled analgesic pumps exported to Europe must simultaneously meet the following requirements:
1) Mechanical performance requirements of YY 0451-2023
2) Drug compatibility study requirements of ISO 28620
3) Add accelerated stability test of drug solution (3 months at 40°C/75%RH)

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

YY 1145-2003 in English
Artificial heart-lung machine terminology
2003-06-17 -

YY 0272-2023 in English
Dentistry-Zinc oxide/eugenol cements and zine oxide/non-eugenol cements
2023-06-20 -

YY/T 0063-1991 in English
Medical diagnostic X-ray tube assembly focus characteristics
1991-11-21 -

YY 0059-1991 in English
dental handpiece
1991-11-21 -

YY 0167-1994 in English
Medical thread
1994-07-02 -

YY 9706.230-2023 in English
Medical electrical equipment-Part 2-30:Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
2023-01-13 -

YY/T 0106-1993 in English
Medical Diagnostic X-ray machine general technical conditions
1993-02-10 -

YY 0188-1993 in English
hip prosthesis
1993-07-19 -

YY 0141-1993 in English
Test methods for suspended particles in clean rooms and clean areas of the pharmaceutical industry
1993-10-18 -

YY 0285.1-2017 in English
Intravascular catheters―Sterile and single-use catheters―Part 1:General requirements
2017-07-17