YY/T 0010-2020 in English
VALIDParticular specifications for dental intra-oral X-ray equipment
- Issued on:2020-06-30
- Implemented on:2021-06-01
- File Format:PDF
- Delivery:Via email within 1~3 business days
$214.00
本标准规定了口内成像牙科X射线机的术语和定义、系统构成、要求和试验方法。本标准适用于使用口腔内X射线影像接收器,用于拍摄牙齿X射线照片的X射线机。本标准不适用于使用口腔外X射线影像接收器成像的X射线机。
Introduction
Background of Standard Update and Technological Evolution
This standard replaces the YY/T 0010-2008 version. The main technological evolution is reflected in:
| Update dimensions | 2008 version | 2020 version |
|---|---|---|
| Standard name | Oral X-ray machine | Intraoral imaging dental X-ray machine |
| Equipment type | Handheld type is not included | Requirements for handheld devices are added |
| Image receiver | Electronic receiver is not specified | Performance indicators of electronic image receivers are added |
| Focus skin distance | No lower limit specified | ≥20cm mandatory requirement |
Analysis of core technical requirements
1. Equipment classification and composition
According to the structure, it is divided into four types: fixed, mobile, portable, and handheld. The basic components include:
- X-ray generating device (including high-voltage generator)
- Telescopic arm system (horizontal rotation ≥40°)
- Electronic image receiver (optional component)
2. Key performance indicators
| Parameter category | Technical requirements | Test method |
|---|---|---|
| X-ray tube voltage | ≥60kV, bin increment≤5kV | GB9706.3-2000 50.104.1 |
| Loading time | Binned according to R10/R20 priority number system | Deviation<±(10%+1ms) |
| Image uniformity | R/Vm≤2.5% | 5 sampling point grayscale analysis |
Special requirements for electronic image receivers
Test case: When using a 6mm pure aluminum attenuation phantom:
- A 0.05mm lead test card is required for line pair resolution testing
- The IQF index needs to be calculated for low contrast resolution
- The image uniformity sampling area is 10%×10%
Implementation recommendations
Manufacturers need to pay special attention to:
- Handheld devices must provide a Radiation Safety Manual
- Electronic receiver products need to add Three image performance tests
- The drop test height is adjusted from 5cm to 50cm hardwood board
It is recommended to establish a Loading factor combination verification process to ensure compliance with the overload protection requirements of 5.3.5.

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

YY/T 0011-2007 in English
X-ray radiographic cassette
2007-01-31 -

YY/T 1959-2025 in English
Specialized technical specifications for digital X-ray imaging systems in the oral cavity
2025-06-18 -

YY/T 0707-2020 in English
Particular specifications for mobile radiography X-ray equipment
2020-09-27 -

YY/T 1944-2024 in English
Special technical requirements for medical X-ray high voltage generators
2024-09-29 -

YY/T 0590.1-2018 in English
Medical electrical equipment—Characteristics of digital X-ray imaging devices—Part 1:Determination of the detective quantum efficiency—Detectors used in radiographic imaging
2018-11-07 -

YY/T 0063-2007 in English
Medical electrical equipment—X-ray tube assemblies for medical diagnosis—Characteristics of focal spots
2007-01-31