Sign In |Help & Support
ALL SECTORS
  • ALL SECTORS
  • GB(National Standard)
  • CB(Shipping)
  • CECS(Engineering Construction)
  • CJ(Urban Construction)
  • CY(News and Publication)
  • DB(Provincial Standard)
  • DL(Electricity & Power)
  • DZ(Geology & Mineralogy)
  • FZ(Spinning & Textile)
  • GA(Public Security)
  • HB(Aviation)
  • HG(Chemical Industry)
  • HJ(Environmental Protection)
  • JB(Machinery)
  • JC(Building Materials)
  • JG(Building & Construction)
  • JJ(Metering)
  • JT(Highway & Transportation)
  • LY(Forestry)
  • MT(Coal)
  • NB(Energy)
  • NY(Agriculture)
  • QB(Light Industry)
  • QC(Automobile & Vehicle)
  • QJ(Aerospace)
  • SH(Petrochemical)
  • SJ(Electronics)
  • SL(Water Resources)
  • SN(Commodity Inspection)
  • SY(Oil & Gas)
  • TB(Railway & Train)
  • YB(Ferrous Metallurgy)
  • YC(Tobacco)
  • YD(Telecommunication)
  • YY(Medical Device)
Database: 365,228(8 Aug 2026)
femoral head fixation method method total hip joint prostheses total hip prostheses stemmed femoral components introduction standard background head fixation stability part gas analysis general quality assurance aspects power safety signs
YY/T 0809.13-2020 in English

YY/T 0809.13-2020 in English

VALID

Implants for surgery—Partial and total hip joint prostheses—Part 13:Determination of resistance to torque of head fixation of stemmed femoral components

  • Issued on:2020-02-21
  • Implemented on:2021-01-01
  • File Format:PDF
  • Delivery:Via email within 1~3 business days
Price(USD): $140.00
$136.00
Standard No: YY/T 0809.13-2020
Document status: VALID
Title in English: Implants for surgery—Partial and total hip joint prostheses—Part 13:Determination of resistance to torque of head fixation of stemmed femoral components
Title in Chinese: 外科植入物 部分和全髋关节假体 第13部分:带柄股骨部件头部固定抗扭转力矩的测定
Language: English
File Format: Electronic (PDF)
Delivery: Via email within 1~3 business days
Issued on: 2020-02-21
Implemented on: 2021-01-01
Professional Classification: YY-Pharmaceutics
Related Keywords: femoral head fixation method method
total hip joint prostheses
total hip prostheses
stemmed femoral components introduction standard background
head fixation
Related Topics: femoral torque
Surgical Implant Partial and Total Hip Prostheses

本标准规定了在特定的实验室条件下测定使髋关节假体股骨球头的固定发生松脱(在髋关节假体中球头预期相对于颈部不能旋转)所需扭矩的方法。
本标准适用于部分或全髋关节假体的股骨部件,其中股骨球头和股骨颈/柄通过圆锥锁定或其他方式固定在一起,其中股骨球头和锥体是独立的部件,它们由金属或非金属材料制成。
本标准不包括检查试样的方法;试样检查应由测试实验室和试样提供方协商。


Introduction

Standard Background and Scope of Application

This standard is modified to adopt the ISO 7206-13:2016 international standard, and establishes a test method for the anti-torsion performance of the head fixation of the stemmed femoral component in partial and total hip prostheses. It mainly solves the problem of loose connection between the femoral ball head and the cone in clinical practice, and is applicable to implant components made of metal/non-metal materials.


Core Test Principle

Through a static torque loading system, apply an increasing torque to the assembly at a rate of 20 N·cm/s±4 N·cm/s or 0.05°/s±0.01°/s until:
1) the peak value of the torque curve decreases
2) the limit value of 50 N·m is reached
3) an angular displacement of 20° occurs

Key ParametersISO 7206-13:2016YY/T 0809.13-2020
Reference StandardsISO 7500-1GB/T 16825.1
Specimen fixing methods3 typesOptimize the order of diagrams
Trademark requirementsLabeling allowedDelete trademark name

Technical requirements for test equipment

A testing machine that complies with GB/T 16825.1 must be equipped with:

  • Axial pressure loading accuracy ≤1%
  • Torque measurement accuracy ≤1%
  • Angular displacement monitoring tolerance ±0.5°

The cone unit must use the same material, process and size specifications as those used in clinical practice.


Specimen preparation specifications

6.3 Femoral head fixation method

Method 1: Processing parallel planes (diameter range 30%-70%)
Method 2: Epoxy resin bonding (covering ≥140° spherical surface)
Method 3: Casting resin embedding

Typical application case: For a 28mm ball head, the recommended plane spacing is 19.0mm±0.5mm, and the remaining wall thickness is ≥2.0mm.


Implementation recommendations

  1. Before testing, the 5-DOF constraints of the fixture system must be verified (only axial rotation is restricted)
  2. It is recommended to use a dynamic torque sensor to monitor the angle-torque curve in real time
  3. After testing, clinical reuse of components is prohibited to avoid changes in material properties affecting secondary testing

Sample only — not a preview of YY/T 0809.13-2020