YY/T 1117-2024 in English
VALIDPlaster bandage
- Issued on:2024-07-08
- Implemented on:2025-07-20
- File Format:PDF
- Delivery:Via email within 1~3 business days
$281.00
| Standard No: | YY/T 1117-2024 |
| Document status: | VALID |
| Title in English: | Plaster bandage |
| Title in Chinese: | 石膏绷带 |
| Language: | English |
| File Format: | Electronic (PDF) |
| Delivery: | Via email within 1~3 business days |
| Issued on: | 2024-07-08 |
| Implemented on: | 2025-07-20 |
| ICS Classification: | 11.120.20-Medical materials |
| Chinese Classification: | C48-Medical sanitary products |
| Professional Classification: | YY-Pharmaceutics |
| Related Keywords: | plaster bandage introduction background
powder type viscose type viscose type ≤15s viscose type uses |
Introduction
Background of Standard Revision and Technological Evolution
The 2024 version of the standard integrates the two independent standards YY/T1117-2001 (powder type) and YY/T1118-2001 (viscosity type) for the first time. The main technological evolution is reflected in:
| Technical Dimensions | 2001 Version | 2024 Version |
|---|---|---|
| Mechanical Properties | Not Specified | Three new indicators including radial compressive strength and peak failure load |
| Curing Control | Time requirement only | Added the restriction of curing temperature ≤39℃ |
| Material composition | Calcium form is not specified | It is stipulated that viscose type uses dihydrate/hemihydrate calcium sulfate |
| Test methods | 6 items | Expanded to 13 items, with a new radial compression test device |
Analysis of key performance indicators
1. Mechanical performance requirements
The newly added radial compression strength indicator is calculated by the formula S=F/d. When testing, please note:
- The specimen preparation needs to be wrapped according to the standard process of 6.5.3
- The transition area should avoid the high tensile stress area
- The test speed is strictly controlled at 20mm/min
2. Curing characteristics control
Dual requirement system:
Powder type: Curing time 4-15min, must not be soft within 24h after curing
Viscose type: Curing time 2-15min, temperature peak ≤39℃ (avoid scalding risk)
Clinical application implementation recommendations
1. Specification selection guide
According to the specification dimensions in Table 1 of the standard, commonly used clinical models are:
| Width (mm) | Applicable parts | Allowance deviation |
|---|---|---|
| 75/100 | Adult limbs | ±2mm |
| 60 | Children/small parts | ±2mm |
| 150 | Trunk fixed | ±3mm |
2. Key points of quality control
- Immersion time: viscose type ≤15s (4.6 times stricter than powder type)
- Gauze density: warp yarn ≥9 yarns/cm (viscose type has 1 less yarn than powder type)
- Chemical properties: calcium sulfate hemihydrate content ≥88% (unified requirements for two types of products)
Impact analysis of standard implementation
Main challenges for manufacturers:
- The radial compression strength test equipment needs to be upgraded (universal testing machine + special fixture)
- The production line needs to add a curing temperature monitoring system
- The adhesive formula needs to be re-verified (clarify the calcium form requirements)
Clinical benefits:
- Mechanical performance indicators improve fixation reliability
- Temperature limits reduce the risk of patient discomfort
- Uniform specifications facilitate horizontal comparison of products

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

YY 0854.2-2011 in English
Performance requirements for pure cotton nonwoven surgical compresses—Part 2:Finished compress
2011-12-31 -

YY/T 1947-2025 in English
Recombinant collagen dressing
2025-02-26 -

YY/T 1938-2024 in English
Medical Sodium Hyaluronate Dressing
2024-09-29 -

YY/T 0729.1-2009 in English
Test methods for boning properties of tissue adhesives—Part 1:Strength in lap-shear by tension loading
2009-06-16