GB/T 32710.10-2016 in English
VALIDSafety requirements for environmental testing and conditioning equipment-Part 10:Electric gravity convection ovens and forced ventilation ovens
- Issued on:2016-06-14
- Implemented on:2017-01-01
- File Format:PDF
- Delivery:Via email within 5 business days
Price(USD):
$280.00
$272.00
$272.00
《GB/T 32710.10-2016环境试验仪器及设备安全规范 第10部分:电热干燥箱及电热鼓风干燥箱》由TC526(全国实验室仪器及设备标准化技术委员会)归口,主管部门为中国机械工业联合会。
Introduction
Analysis of the core content of the standard
| Safety dimension | Mandatory requirements | Test methods |
|---|---|---|
| Protection against electric shock | Contact current ≤0.5mA (normal)/3.5mA (fault) | Chapter 6 of GB 4793.1-2007 |
| Surface temperature | ≤105℃ (single fault condition) | Multi-point detection with infrared thermometer |
| Over-temperature protection | Double thermal circuit breakers (at least 1 non-automatic reset) | 4.4.2.9 Fault test |
Key technical requirements
1. Temperature control system
Standard requirementsElectric drying oven must be equipped with an over-temperature protection device independent of the main control system:
- The main heating element must be equipped with a double thermal circuit breaker, at least 1 of which is a non-automatic reset type
- The operating temperature of the protection device should be at least 50°C lower than the ignition point of the material
- Walk-in equipment must be equipped with an internal emergency unlocking device
2. Mechanical safety design
Special requirements for Electric blast drying oven:
- The door lock mechanism needs to prevent finger pinching (gap ≤4mm)
- The circulation fan is interlocked with the door, and the heating power is automatically cut off when the door is opened
- The glass observation window needs to be tempered and marked with an anti-scalding mark
Implementation suggestions
Key points for equipment selection
- Confirm that the protection level (IP code) marked on the equipment nameplate meets the use environment
- Check whether the over-temperature protection device has an independent temperature sensor
- IP21 and above protection models should be used in humid places
Maintenance and management specifications
- Test the over-temperature protection device function every month (simulate over-temperature conditions)
- Check the sealing of the door seal every quarter (heat leakage ≤5%)
- Calibrate the temperature sensor annually (error within ±1℃)
Sample only — not a preview of GB/T 32710.10-2016

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.