SN/T 5151-2019 in English
VALIDDetermination of seven benzophenones in sunscreen cosmetics by high performance liquid chromatography
- Issued on:2019-10-25
- Implemented on:2020-05-01
- File Format:PDF
- Delivery:Via email within 1~3 business days
$175.00
Introduction
Technical background and significance of the standard
As a mandatory inspection procedure issued by the General Administration of Customs in 2019, this standard establishes a systematic detection method for 7 potential photosensitizing benzophenone derivatives in sunscreen cosmetics. As common ultraviolet absorbers, excessive use of benzophenones may cause skin allergic reactions. EU Cosmetics Regulation (EC) No 1223/2009 has set limit requirements for some substances.
Core detection principle
Baseline separation of 7 structurally similar substances is achieved by using a high performance liquid chromatograph combined with a UV detector through a gradient elution program. The technical routes include:
- Tetrahydrofuran-methanol mixed solvent extraction
- C18 reverse phase chromatography column separation (250mm×4.6mm, 5μm)
- External standard method quantification (0.5-20μg/mL linear range)
- LC-MS/MS positive result confirmation
Comparison of key operating parameters
| Analyte | CAS number | Lower limit of determination (mg/g) | Recovery range | Characteristic ion pairs (m/z) |
|---|---|---|---|---|
| Benzophenone-1 | 131-56-6 | 0.1 | ||
| Benzophenone-3 | 131-57-7 | 0.2 | 87.1-94.5% | 229.4/151.5 |
| Benzophenone-12 | 1843-05-6 | 0.4 | 84.9-90.3% | 327.3/137.6 |
Method Validation Data
Collaborative verification by three laboratories showed that:
- Intra-day precision RSD≤5.8% (n=6)
- Inter-day precision RSD≤8.5% (n=3)
- Matrix effect influence rate <15%
In the actual detection of a certain brand of sunscreen, the content of dibenzophenone-3 was detected at 1.28 mg/g, and the spike recovery rate reached 91.4%.
Recommendations for the implementation of the standard
1. Pretreatment optimization: For physical sunscreen products containing titanium dioxide, it is recommended to add 10 minutes of centrifugation (12000rpm) to remove particle interference
2. Quality control points: Each batch of testing should include blank controls, matrix spikes and parallel samples. It is recommended to use isotope internal standards (such as dibenzophenone-d10) to correct for matrix effects
3. Equipment maintenance: The C18 column needs to be rinsed with 95% acetonitrile for 30 minutes after every 100 injections to keep the column efficiency stable

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

SN 0283-1994 in English
\"Method for Inspection of Dichlorodipyridinol Residues in Exported Poultry Meat\"
1994-02-18 -

SN/T 0244-1994 in English
\"General Rules for Inspection and Certification of Imported Mechanical and Electrical Instruments\"
1993-08-01 -

SN/T 0839.4-1999 in English
Industrial glycerols for import and export - Determination of water content - Karl Fischer method
1999-12-30 -

SN/T 4497-2016 in English
Specification on the assessment of severity levels of potential injury for import and export consumer products
2016-06-28 -

SN/T 1164.1-2011 in English
Quarantine Protocol for Infectious Bovine Rhinotracheitis
2011-05-31 -

SN/T 1171-2011 in English
Quarantine Protocol for Caprine Arthritis-encephalitis and Maedi-visna of Sheep
2011-02-25