Introduction
Analysis of the core content of the standard
This standard for the first time incorporates Fourier transform near-infrared spectrometer and grating scanning fiber spectrometer into the raw cotton regain detection system, and specifies the technical parameter requirements of the two instruments:
| Instrument type | Wavelength range | Resolution | Signal-to-noise ratio | Modeling method |
|---|---|---|---|---|
| Fourier transform type | 4000-10000 cm⁻¹ | 8 cm⁻¹ | 10000:1 | Partial least squares method |
| Raster scanning type | 1000-1800 nm | 10 nm | 30000:1 | Partial least squares method |
Technical principles and innovations
Rapid detection is achieved by establishing a mathematical model of the raw cotton spectrum and the values measured by the oven method, utilizing the combined frequency absorption band characteristics of water molecules in the near-infrared region. Key technological breakthroughs include:
- Savitzky-Gloay smoothing and first-order derivative preprocessing are used to eliminate background interference
- Model transfer technology ensures consistency of results between different instruments (absolute error ≤ 0.5%)
- Three-dimensional sampling design (front/back/inner layer) reduces the impact of fiber unevenness
Key points of operating specifications
Typical application cases
In a customs laboratory's test of batches of U.S. cotton, the near-infrared method measured a regain of 8.2%±0.3%, which was consistent with the result of 8.0% by the oven method. The testing time for a single sample was shortened from 4 hours to 5 minutes.
Abnormal handling process
When the difference between two measurements is >0.5% or the result is outside the range of 6-15%, you need to:
- Check whether the sample contains weeds/foreign fibers
- Resample and test
- If still abnormal, switch to the GB/T 6102.1 oven method
Implementation suggestions
Laboratories should pay attention to the following when introducing this standard:
- Instrument verification:Must pass the accuracy (SEP≤1.0%) and repeatability (≤0.5%) verification in Appendix A
- Model maintenance:Use monitoring samples for daily calibration, and remodeling is required if the deviation exceeds the limit
- Staff training:Focus on mastering the spectral acquisition specifications (sampling bottle pressure, rotation speed, etc.)

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

SN 0283-1994 in English
\"Method for Inspection of Dichlorodipyridinol Residues in Exported Poultry Meat\"
1994-02-18 -

SN/T 2535-2010 in English
Rules for labeling of import and export refrigerants
2010-03-02 -

SN/T 0244-1994 in English
\"General Rules for Inspection and Certification of Imported Mechanical and Electrical Instruments\"
1993-08-01 -

SN/T 1279-2011 in English
Treatment Codes for Entry-exit Special Products
2011-05-31