Sign In |Help & Support
ALL SECTORS
  • ALL SECTORS
  • GB(National Standard)
  • CB(Shipping)
  • CECS(Engineering Construction)
  • CJ(Urban Construction)
  • CY(News and Publication)
  • DB(Provincial Standard)
  • DL(Electricity & Power)
  • DZ(Geology & Mineralogy)
  • FZ(Spinning & Textile)
  • GA(Public Security)
  • HB(Aviation)
  • HG(Chemical Industry)
  • HJ(Environmental Protection)
  • JB(Machinery)
  • JC(Building Materials)
  • JG(Building & Construction)
  • JJ(Metering)
  • JT(Highway & Transportation)
  • LY(Forestry)
  • MT(Coal)
  • NB(Energy)
  • NY(Agriculture)
  • QB(Light Industry)
  • QC(Automobile & Vehicle)
  • QJ(Aerospace)
  • SH(Petrochemical)
  • SJ(Electronics)
  • SL(Water Resources)
  • SN(Commodity Inspection)
  • SY(Oil & Gas)
  • TB(Railway & Train)
  • YB(Ferrous Metallurgy)
  • YC(Tobacco)
  • YD(Telecommunication)
  • YY(Medical Device)
Database: 365,228(8 Aug 2026)
diagnostic display breast medical diagnostic display device introduction analysis diagnostic display test basis resolution ≥1600×1200 ≥2048×2560 sj/t11292-2016 breast diagnosis display display area edge integrity error technical requirements technical dimensions requirements low-production lithium batteries capital construction items
T/CAME 33-2021 in English

T/CAME 33-2021 in English

VALID

Medical diagnostic display device

  • Issued on:2021-06-30
  • Implemented on:2021-06-30
  • File Format:PDF
  • Delivery:Via email within 1~3 business days
Price(USD): $100.00
$97.00

本文件规定了医用诊断显示器的技术要求以及试验方法。
本文件适用于普通医用诊断显示器和乳腺医用诊断显示器。


Introduction

Analysis of the core technical requirements of the standard

Parameters General diagnostic display Breast diagnostic display Test basis
Resolution ≥1600×1200 ≥2048×2560 SJ/T11292-2016 5.6.12
Calibrated luminance (cd/m²) ≥350 ≥450 YY/T0910.1-2013 7.4.1
Contrast ratio ≥300 ≥500 YY/T0910.1-2013 7.4.1

Detailed explanation of key technical indicators

Brightness uniformity requirements

The standard adopts the nine-point measurement method (screen size ≥23 inches), and the calculation formula is: Lu=100-200×(Lhighest-Llowest)/(Lhighest+Llowest). The breast diagnosis display requires ≥85%, which is 10 percentage points higher than the ordinary type.

TG18-OIQ Test Specification

Contains 8 visual inspection items, with special requirements: Breast Diagnosis Display must fully display the letters "QUALITYCONTROL" in the black area, while the ordinary type is allowed to lack the last 3 letters.


Clinical Application Recommendations

Equipment Selection Guide

1. Radiology CT/MRI diagnosis should choose ordinary diagnostic monitors, which can meet the brightness of 350cd/m²
2. Mammography examination must be equipped with a dedicated breast diagnostic monitor, ensuring a high brightness of 450cd/m²
3. PACS reading workstations are recommended to have dual screens, with the main screen resolution not less than 2048×2560


Key Points for Implementation of Standards

1. A color analyzer should be used for grayscale chromaticity detection every quarter (Δu'v'≤0.01)
2. When detecting brightness response error, the tolerance of the breast monitor is only 5%, and professional calibration equipment is required
3. The grid test requires that the display area edge integrity error is less than 1%

Sample only — not a preview of T/CAME 33-2021