YY 0777-2023 in English
VALIDRadio frequency hyperthermia equipment
- Issued on:2023-01-13
- Implemented on:2026-01-15
- File Format:PDF
- Delivery:Via email within 1~3 business days
$156.00
| Standard No: | YY 0777-2023 |
| Document status: | VALID |
| Title in English: | Radio frequency hyperthermia equipment |
| Title in Chinese: | 射频热疗设备 |
| Language: | English |
| File Format: | Electronic (PDF) |
| Delivery: | Via email within 1~3 business days |
| Issued on: | 2023-01-13 |
| Implemented on: | 2026-01-15 |
| Superseding: |
YY 0777-2010 Radio frequency hyperthermia equipment
|
| Professional Classification: | YY-Pharmaceutics |
| Related Keywords: | radio frequency hyperthermia equipment introduction background
operating frequency range new frequency stability requirement frequency meter multi-point monitoring temperature control accuracy ±1℃ ±0.5℃ accuracy |
| Related Topics: | RF equipment use
RF equipment RF equipment therapy hyperthermia treatment+ frequency Frequency+ bluetooth radio frequency device RF heater RF instrument design Medical RF equipment accessories Radiofrequency hyperthermia equipment industry |
本文件规定了射频热疗设备的要求和试验方法。
本文件适用于射频热疗设备。
本文件不适用于射频消融设备。
Introduction
Background of Standard Revision and Technology Evolution
This standard, as an alternative version of YY 0777-2010, reflects the development achievements of radiofrequency thermal therapy technology over the past decade. It mainly deletes non-core clauses such as appearance inspection and composition structure, and strengthens key technical indicators such as output power accuracy (±20%) and temperature control (±0.5℃).
Comparison of core technical requirements
| Technical parameters | 2010 version requirements | 2023 version requirements | Range of change |
|---|---|---|---|
| Output power error | Not specified | ≤±20% | New quantitative indicators |
| Temperature measurement channels | Unspecified number | ≥5 channels | Add multi-point monitoring |
| Temperature control accuracy | ±1℃ | ±0.5℃ | Accuracy improved by 50% |
Details of key technical specifications
1. Electromagnetic parameter requirements
The operating frequency range remains 3MHz-120MHz, but a new frequency stability requirement is added (nominal value ±2%). The test needs to be measured at rated power using a frequency meter (accuracy ±0.5%).
2. Temperature control system
Use fiber optic thermometer (accuracy ±0.1℃) as the standard reference, and require the equipment to:
- Equipped with automatic temperature control mode
- Settings above 45℃ require special operation to unlock
- Temperature fluctuation curve recording interval ≤ 1 minute
Safety protection mechanism
Typical safety design case: When the temperature measurement channel is abnormal or the communication is interrupted, the system needs to alarm at a sound pressure level of 60dB(A) and automatically cut off the RF output, and the visual warning needs to continue until manual reset. The emergency stop device must meet the requirement of single action to cut off all power sources.
Implementation suggestions
- Upgrade the power calibration system to ensure that the power output error within the full range is ≤±15% (reserve a 5% margin)
- Use a distributed temperature sensing network, and it is recommended to configure 8-10 temperature measurement points (exceeding the minimum standard requirement)
- Electromagnetic compatibility design must meet the dual requirements of GB 9706.1 and GB 9706.203-2020

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

YY 1145-2003 in English
Artificial heart-lung machine terminology
2003-06-17 -

YY/T 0290.5-2023 in English
Ophthalmic optics - Intraocular lenses - Part 5: Biocompatibility
2023-09-05 -

YY 0167-1994 in English
Medical thread
1994-07-02 -

YY 0059-1991 in English
dental handpiece
1991-11-21 -

YY/T 0085.1-1992 in English
Pulse vacuum pressure steam sterilizer
1992-01-21 -

YY/T 0063-1991 in English
Medical diagnostic X-ray tube assembly focus characteristics
1991-11-21 -

YY/T 1892-2024 in English
Break point cluster region - Abelson leukemia virus(BCR-ABL)fusiongene testing kit
2024-02-07 -

YY/T 0162-1994 in English
Medical Ultrasound Equipment Grade Series Part 1 Type B Ultrasound Diagnostic Equipment
1994-05-13 -

YY 0450.2-2003 in English
Accessory devices for sterile single-use intravascular catheters - Part 2: Obturators for over-needle peripheral catheters
2003-02-09