YY/T 0905.1-2016 in English
SUPERSEDEDDentistry-Plant area equipment-Part 1: Suction systems
- Issued on:2016-03-23
- Implemented on:2017-01-01
- File Format:PDF
- Delivery:Via email within 5 business days
$301.00
| Standard No: | YY/T 0905.1-2016 |
| Document status: | SUPERSEDED |
| Superseded by: |
YY/T 0629-2021 Dentistry—Central suction source equipment
|
| Superseded on: | 2022-04-01 |
| Title in English: | Dentistry-Plant area equipment-Part 1: Suction systems |
| Title in Chinese: | 牙科学场地设备 第1部分:吸引系统 |
| Language: | English |
| File Format: | Electronic (PDF) |
| Delivery: | Via email within 5 business days |
| Issued on: | 2016-03-23 |
| Implemented on: | 2017-01-01 |
| ICS Classification: | 11.060.20-Dental equipment |
| Professional Classification: | YY-Pharmaceutics |
| Related Keywords: | suction systems introduction standard framework
suction connection points suction ports suction machine core term definition termdefinitionkey indicators suction machinedevice |
| Related Topics: | Chemical system equipment
site scientific equipment scientific equipment Equipment Science Equipment Science Site equipment Dentistry attract yy/t 0905.1-2016 |
本标准适用场地(场地指牙科治疗的设备或设施所安装的位置)中的牙科吸引设备。吸引设备包括:吸引机、银汞合金分离器、过滤器、阀门、管道、管道配件。这些设备用于YY/T 0629―2008中规定的牙科设备的吸引源。
本标准规定了牙科吸引设备的性能、试验方法和排气要求,且仅限于吸引管连接点之后的吸引系统的性能。
Introduction
Standard Framework and Development Background
This standard revision adopts ISO/TS 22595-1:2006 international standard and combines the technical transformation with the Chinese medical device regulatory requirements. The main adjustments include: the pressure unit is unified as kPa, the reference standard is replaced with the current domestic version, and the specific requirements of 11 test methods are supplemented.
Core term definition
| Term | Definition | Key indicators |
|---|---|---|
| Suction machine | Device that generates negative pressure through electric drive | Maximum vacuum degree ≤25kPa |
| Amalgam separator | Device for intercepting heavy metals in wastewater | In compliance with YY 0835-2011 |
| Bacteria filter | Device for removing microorganisms from the air | Filtration accuracy 0.01μm |
Key technical requirements
5.2.17 Safety valve performance
The system must be equipped with a safety valve. When all suction ports are closed, the vacuum value must not exceed 25kPa. The test method is: run the system after closing all suction connection points, and measure the pressure value with a precision vacuum gauge.
5.3.1 Flow rate standard
- High capacity system: each interface ≥250L/min
- Medium capacity system: each interface ≥90L/min
During the test, the suction machine characteristic curve provided by the manufacturer must be combined, and the flow rate must be converted by measuring the pressure point data of the section pipe.
Implementation points
Installation specifications
The exhaust pipe must be an independent pipe, and the air outlet should be:
1. It should be installed on the roof of the building first
2. It should be ≥3m away from doors and windows
3. It should be equipped with a backflow prevention device
Maintenance suggestions
The bacterial filter needs to be replaced regularly, and maintenance should be performed immediately when the differential pressure gauge reading exceeds 5kPa. The separation efficiency of the silver-mercury alloy separator needs to be tested monthly according to the requirements of YY 0835.
Comparison of old and new standards
| Project | ISO 22595-1:2006 | YY/T 0905.1-2016 |
|---|---|---|
| Pressure unit | mbar | kPa (equivalent conversion) |
| Noise standard | ISO 374 series | GB/T 4980-2003 |
| Electrical safety | IEC 60364 | GB 16895.24 |