YY/T 1309-2016 in English
VALIDWasher-disinfectors-Requirement and test of ultrasonic wash
- Issued on:2016-03-23
- Implemented on:2017-01-01
- File Format:PDF
- Delivery:Via email within 1~3 business days
$127.00
本标准适用于具有自动超声清洗功能且符合YY/T 0734.1-2009和YY/T 0734.2-2009要求的清洗消毒器。
本标准规定了超声清洗的专用要求,并与YY/T 0734.1-2009、YY/T 0734.2-2009中的规定合并使用。
Introduction
Interpretation of the core content of the standard
As a special supplement to the YY/T 0734 series of standards, this standard proposes 21 technical indicators for the ultrasonic cleaning module of the cleaner-disinfector. The main specifications are:
| Technical dimensions | Requirements of this standard | Basic standard requirements |
|---|---|---|
| Frequency accuracy | ±2kHz deviation | Not specified |
| Power density | 0.3-0.55W/cm² | No relevant clauses |
| Dielectric volume ratio | 10-40W/L | No relevant clauses |
Analysis of key technical requirements
4.1 Ultrasonic cavity design
The standard stipulates that the cavity must meet: 120℃ high temperature tolerance, ΔT≤35℃ temperature rise limit, and mandatory observation window/liquid level indicator. In practical applications, it is recommended to use sintering adhesive bonding process to ensure the reliability of transducer connection.
4.5-4.7 Power parameters
The power-to-volume ratio is calculated using the formula PH=Pp/V, where:
- Input power measurement requires the use of a power meter with an accuracy of ≥1%
- The transducer efficiency η should be provided in writing by the manufacturer
Implementation recommendations
Application case in medical institutions
When upgrading the fully-automatic cleaning and disinfection device, the disinfection supply center of a tertiary hospital selected equipment with a power-to-volume ratio of 25W/L in accordance with Clause 4.6 of this standard, which increased the qualified rate of protein residue detection in surgical instruments by 32%.
Key points for manufacturers’ compliance
- A frequency-power correspondence curve must be provided
- Glass instrument compatibility testing must include 20 cycle tests
- Noise testing must be measured in four directions at a distance of 1m

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

YY 0249-1996 in English
Aluminum tube for medicinal ointment
1996-12-30 -

YY/T 0063-1991 in English
Medical diagnostic X-ray tube assembly focus characteristics
1991-11-21 -

YY/T 0085.2-1992 in English
Forevacuum pressure steam sterilizer
1992-01-21 -

YY/T 0106-1993 in English
Medical Diagnostic X-ray machine general technical conditions
1993-02-10 -

YY 0141-1993 in English
Test methods for suspended particles in clean rooms and clean areas of the pharmaceutical industry
1993-10-18 -

YY 0233-1995 in English
Vertical ampoule production line
1996-12-22 -

YY/T 0851-2023 in English
Medical thrombosis prophylaxis hosiery
2023-09-05 -

YY/T 0085.1-1992 in English
Pulse vacuum pressure steam sterilizer
1992-01-21 -

YY 1145-2003 in English
Artificial heart-lung machine terminology
2003-06-17 -

YY 0167-1994 in English
Medical thread
1994-07-02