Sign In |Help & Support
ALL SECTORS
  • ALL SECTORS
  • GB(National Standard)
  • CB(Shipping)
  • CECS(Engineering Construction)
  • CJ(Urban Construction)
  • CY(News and Publication)
  • DB(Provincial Standard)
  • DL(Electricity & Power)
  • DZ(Geology & Mineralogy)
  • FZ(Spinning & Textile)
  • GA(Public Security)
  • HB(Aviation)
  • HG(Chemical Industry)
  • HJ(Environmental Protection)
  • JB(Machinery)
  • JC(Building Materials)
  • JG(Building & Construction)
  • JJ(Metering)
  • JT(Highway & Transportation)
  • LY(Forestry)
  • MT(Coal)
  • NB(Energy)
  • NY(Agriculture)
  • QB(Light Industry)
  • QC(Automobile & Vehicle)
  • QJ(Aerospace)
  • SH(Petrochemical)
  • SJ(Electronics)
  • SL(Water Resources)
  • SN(Commodity Inspection)
  • SY(Oil & Gas)
  • TB(Railway & Train)
  • YB(Ferrous Metallurgy)
  • YC(Tobacco)
  • YD(Telecommunication)
  • YY(Medical Device)
Database: 365,228(8 Aug 2026)
warshot torpedo plastic injection mold parts pollution resistance test
DL/T 1954-2018 in English

DL/T 1954-2018 in English

VALID

Calibration specification of partial discharge detector based on transient earth voltage

  • Issued on:2018-12-25
  • Implemented on:2019-05-01
  • File Format:PDF
  • Delivery:Via email within 1~3 business days
Price(USD): $280.00
$272.00


Introduction

Interpretation of the core content of the standard

This standard specifies the metrological characteristic requirements and calibration methods of the transient ground voltage partial discharge detector, and is applicable to the calibration of equipment in the 3MHz-100MHz frequency band.


Analysis of key technical indicators

Metering characteristicsTechnical parametersAllowable error
Frequency band range3MHz-100MHz≤±10%
Amplitude measurementdBmV unit≤±3dBmV
Pulse repetition rate≥10kHz≤±5%
SensitivityNoise reference 2 timesIndication ratio≥1

Calibration Implementation Points

Typical Calibration Process

  1. Safety Performance Check: Including 20MΩ insulation resistance test and 2kV/1min withstand voltage test
  2. Frequency Characteristics Calibration: Using a sine wave signal generator, measure the -6dBmV cutoff point
  3. Amplitude Linearity Calibration: Measure 5 points every 6dBmV from 0dBmV
  4. Pulse Repetition Rate Verification: Compare the difference between the standard generator and the actual measurement value

Measurement Uncertainty Analysis

Take the 40dBmV calibration point as an example:

  • Uncertainty introduced by the standard: 0.0577dBmV (Class B)
  • Uncertainty caused by resolution: 0.5773dBmV (Class B)
  • Repeatability uncertainty: 0.4216dBmV (Class A)
  • Combined expanded uncertainty: 1.4dBmV (k=2)

Standard evolution and implementation suggestions

Technical development background

This standard was first released (2018), filling the gap in the calibration standard for TEV method detectors and forming a supporting system with DL/T846.10-2016.

Application Recommendations

  • Calibration cycle: 1 year is recommended, the cycle should be shortened in harsh environments
  • Equipment selection: give priority to detectors with automatic calibration function
  • Field application: pay attention to the control of ambient temperature (15-25℃) and humidity (20-80%)

Sample only — not a preview of DL/T 1954-2018
Page: 1 / 0
100%

Loading PDF document...

Error loading PDF. Please make sure the file is valid and try again.

We also recommend

  • DL/T 790.12-2001 in English

    DL/T 790.12-2001 in English

    Distribution Automation Using Distribution Line Carrier Part 1: General Provisions Part 2: Guidelines for Formulation of Specifications

    2001-12-26
  • DL/T 863-2016 in English

    DL/T 863-2016 in English

    Guide for Turbine Start-up Commissioning Test

    2016-08-16