GB/T 19634-2021 in English
VALIDIn vitro diagnostic test systems—General technical requirements for blood-glucose monitoring systems for self-testing
- Issued on:2021-10-11
- Implemented on:2023-05-01
- File Format:PDF
- Delivery:Via email within 1~3 business days
$214.00
《GB/T 19634-2021体外诊断检验系统 自测用血糖监测系统通用技术条件》由TC136(全国医用临床检验实验室和体外诊断系统标准化技术委员会)归口,主管部门为国家药品监督管理局。
Introduction
Analysis of the core revisions to the standard
Compared with the 2005 version, GB/T19634-2021 has mainly made 11 technical revisions, with a focus on strengthening the safety requirements in self-test scenarios:
| Technical Dimensions | Requirements of the 2005 version | Improvements of the 2021 version |
|---|---|---|
| Accuracy Deviation | ±20% | ±15%(≥5.55mmol/L) |
| Hematocrit | Not required | Newly added ±0.55mmol/L deviation limit |
| Electromagnetic compatibility | No requirement | Newly added special requirements of GB/T18268.26 |
Detailed explanation of key performance indicators
1. System accuracy requirements
The new standard adopts segmented deviation control:
- Low concentration range (<5.55mmol/L): absolute deviation ≤±0.83mmol/L
- High concentration range (≥5.55mmol/L): relative deviation ≤±15%
Application case: A certain brand
2. Influence of hematocrit
The standard first establishes the hematocrit-blood glucose concentration matrix test requirements:
| Test dimensions | Technical requirements |
|---|---|
| Hematocrit level | ≥5 gradients |
| Blood glucose concentration | 3 characteristic points (low/medium/high) |
| Allowable deviation | ±0.55mmol/L (low concentration) |
Implementation recommendations
Compliance path for manufacturers
- Establish a three-level calibration system: working calibration product → manufacturer's standing procedure → reference measurement procedure
- Improve electromagnetic compatibility design: must pass the RF immunity test specified in GB/T18268.26
- Optimize test strip batch control: use the Six Sigma method to control the coefficient of variation of film thickness
Precautions for clinical use
- Additional compensation calibration is required in areas with an altitude of >2000 meters
- When the hematocrit is <35%, the plasma conversion mode should be used
- The frequency of quality control testing should not be less than once every 24 hours

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

GB/T 18990-2008 in English
Luteinizing hormone(LH)test strip(Colloidal gold immunochromatographic assay)
2003-03-05 -

GB/T 40672-2021 in English
Clinical laboratory testing—Criteria for acceptable lots of dehydrated Mueller-Hinton agar and broth for antimicrobial susceptibility testing
2021-10-11 -

GB/T 29791.1-2013 in English
In vitro diagnostic medical devices―Information supplied by the manufacturer(labelling)―Part 1:Terms,definitions and general requirements
2013-10-10 -

GB/T 26124-2011 in English
In vitro diagnostic reagent(kit) for clinical chemistry
2011-05-12