Calibration Specification for the Equipment of the Environmental Testing for Temperature and Humidity 环境试验设备温度、湿度校准规范 1 Scope This specification is applicable to the calibration for metrological performance of equipment of the environmental testing for temperature and humidity (hereinafter referred to as environmental testing equipment). Other similar equipment may also be calibrated by reference to this specification. 2 References GB/T 2423.4-1993 Basic Environmental Testing Procedures for Electric and Electronic Products - Test Db: Damp Heat, Cyclic GB 6999-1986 Relative Humidity Tables for Environmental Test JJF 1071-2000 The Rules for Drafting National Calibration Specification While using this specification, current valid editions of such documents shall be applied. 3 Terms 3.1 Temperature Deviation The difference between the displayed temperature average of the environmental testing equipment under stable state and the measured temperature average of the center point in work space. 3.2 Relative Humidity Deviation The difference between the displayed relative humidity average of the environmental testing equipment under stable state and the measured relative humidity average of the center point in work space. 3.3 Nominal Value The parameter specified according to the testing method requirements or that predetermined as required when calibrating the equipment of the environmental testing for temperature and humidity. 4 Metrological Performance See Table 1 for the deviation, uniformity and fluctuation respectively for the temperature and relative humidity. 5 Calibration Conditions 5.1 Environmental Conditions 5.1.1 Temperature: (15~35)℃ Humidity: (30~85)%RH Atmospheric pressure: (86~106)kPa 5.1.2 Loading conditions Generally, calibration is carried out under no-load condition; where required by the user, it may be carried out under loaded condition. However, the loading condition shall be descripted. Table 1 Equipment name Temperature testing equipment Damp heat testing equipment Alternating damp heat equipment Temperature and humidity range (0~100) ℃ (-60~0)℃ (100~200)℃ (200~300) ℃ (10~60)℃ (20~100) %RH (10~60)℃ (20~100) %RH (20~60)℃ (80~100) %RH Temperature deviation ±1.0℃ ±2℃ ±3℃ ±2℃ ±2℃ ±2℃ Humidity deviation - - - +2 %RH -3 ±5%RH +2 %RH -3 Uniformity Temperature 1.0℃ 2℃ 3℃ 1℃ 2℃ 2℃ Humidity - - - 3%RH 5%RH 3%RH Fluctuation Temperature ±0.5℃ ±0.5℃ ±2℃ ±0.5℃ ±1℃ ±1℃ Humidity - - - ±2%RH ±3%RH ±3%RH Notes: 1. The alternating capacity of alternating damp heat equipment shall be inspected. 2. For the temperature and damp heat testing equipment with metrological performance otherwise specified, calibration shall be carried out according to the requirements of relevant technical documents. 5.1.3 Other conditions: the equipment shall be free from strong vibration and corrosive gas at its surrounding, and shall be free from being affected by other cold and heat sources. 5.2 Standard Instrument and Other Equipment 5.2.1 Temperature measurement standard Temperature measurement is composed of temperature sensor (generally, four-wire system platinum thermistor is adopted) and displayer, and the time constant shall be less than 15s. 5.2.2 Humidity measurement standard The following instruments may be adopted for humidity measurement: Digital ventilated psychrometer and gas-pressure meter (the ventilation speed shall be greater than 2.5m/s) Digital hygrometer (only used for humidity field without alternation) Dry and wet bulb thermometers① (applicable to the measurement of relative humidity and uniformity under uniform wind speed) 6 Calibration Items and Methods 6.1 Calibration Items See Table 2 for the calibration items. Note: the manufacture method of ① dry and wet bulb thermometers shall meet the requirements of Appendix C.
Scope
This specification is applicable to the calibration of temperature and humidity measurement performance of environmental test equipment. Other similar equipment can also be calibrated with reference to this specification.
1 Scope
2 References
3 Terms
3.1 Temperature Deviation
3.2 Relative Humidity Deviation
3.3 Nominal Value
4 Metrological Performance
5 Calibration Conditions
5.1 Environmental Conditions
5.2 Standard Instrument and Other Equipment
6 Calibration Items and Methods
6.1 Calibration Items
6.2 Calibration Methods
6.3 Data Processing
7 Calibration Result Expression
8 Recalibration Interval
Appendix A Reference Format for Calibration Record of Environmental Testing Equipment
Appendix B Reference Format for Calibration Result of Environmental Testing Equipment
Appendix C Relative Humidity Measurement with Dry and Wet Bulb Method
Appendix D Uncertainty Analysis for Temperature Deviation Calibration Result of Environmental Testing Equipment
Appendix E Uncertainty Analysis for Relative Humidity Deviation Calibration Result of Environmental Testing Equipment

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

JJF 1353-1990 in English
O.T.N. for Primary Standard of Combustion Energy
1990-01-01 -

JJF 1098-2003 in English
Calibration Specification for Auto-measuring System of Thermocouples and Resistance Thermometers
2003-03-05 -

JJF 1215-1990 in English
O.T.N. of Temperature Primary Standard in the Range from 12.81 to 273.15K
1990-01-01 -

JJF 1049-1995 in English
Calibration of Temperature Sensors Dynamic Reponse
1995-11-07 -

JJF 1370-1995 in English
Operating technical norm of water calorimeter absorbed dose primary standard for γ-rays
1994-12-01 -

JJF 1219-1990 in English
O.T.N.of Temperature Primary Standard in the Range from 273.15 to 903.89K
1990-01-01