JJG 700-2016 in English
VALIDVerification Regulation of Gas Chromatograph
- Issued on:2016-06-27
- Implemented on:2016-12-27
- File Format:PDF
- Delivery:Via email within 1~3 business days
$253.00
1 Scope
This regulation is applicable to initial verification, subsequent verification and in-service inspection of gas chromatographs equipped with thermal conductivity detector (TCD), flame ionization detector (FID), flame photometric detector (FPD), electron capture detector (ECD) and nitrogen phosphorous detector (NPD).
2 Overview
A gas chromatograph (hereinafter referred to as the instrument) is an instrument in which the sample is carried into the chromatographic column by a carrier gas, separated based on the distribution of sample components in the gas phase and stationary phase of chromatographic column as well as the different adsorption coefficients and tested with the detector. Qualitative and quantitative analysis is made according to retention time and response value of the components.
The instrument consists of the gas path system, sample injection system, separation system, temperature control system, detection system and data processing system.
3 Metrological Performance Requirements
Metrological performance of the instrument shall meet the requirements specified in Table 1
... ...
... ...
4 General Technical Requirements
4.1 Appearance
The instrument shall be free of damages affecting its normal operation. The switches, knobs, or buttons shall be operable and controllable normally and the display of indicator light shall be clear and correct. The manufacturer name, model, serial number and date of manufacture shall be indicated on the instrument. Domestic instruments shall be provided with the mark and number of China Metrology Certification.
Introduction i
1 Scope
2 Overview
3 Metrological Performance Requirements
4 General Technical Requirements
4.1 Appearance
4.2 Gas path System
5 Control of Metrological Instruments
5.1 Verification Conditions
5.2 Reference Materials and Equipment Used for Verification
5.3 Verification Items
5.4 Verification Method
5.5 Treatment of Verification Results
5.6 Verification Period
Annex A Correction of Carrier Gas Flow Rate
Annex B Conversion of Molar Fraction of Gas Reference Materials to Concentration
Annex C Verification Record Format
Annex D Inside Page Format for Verification Certificate/Verification Result Notice

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

JJG 496-2016 in English
Verification Regulation of High-voltage Divider at Power Frequency
2016-11-25 -

JJG 795-2016 in English
Withstanding Voltage Testers
2016-11-25 -

JJG 518-2023 in English
Pitot Tubes
2023-03-15 -

JJG 450-2016 in English
Fruits Hardness Testers
2016-11-30 -

JJG 703-2003 in English
Verification Regulation of Electro-optical Distance Meter (EDM Instruments)
2003-09-23 -

JJG 86-2023 in English
Standard Glass Hydrometers
2023-03-15 -

JJG 802-2019 in English
Distortion Meter Calibrators
2019-09-27 -
17-1991.png)
JJG (石油) 17-1991 in English
Dritling Fluid Rotary Viscosineter
1999-12-01 -

JJG 26-2011 in English
Micrometers with Dial Comparator and Indication Snap Gauge
2011-09-14 -

JJG 603-2018 in English
Verification Regulation of Frequency Meters
2018-12-25