VALIDNatural gas-Determination of the hydrocarbon dew point- Cooled mirror automatic method
- Issued on:2020-10-23
- Implemented on:2021-02-01
- File Format:PDF
- Delivery:Via email within 1~3 business days
$156.00
Introduction
Interpretation of the core content of the standard
This standard specifies the method of measuring the hydrocarbon dew point of pipeline natural gas using a cooling mirror automatic detector, with a measurement accuracy of ±0.5℃. It is applicable to single-phase natural gas, but not to gas containing liquid hydrocarbons.
Technical principles and instrument requirements
| Components | Technical requirements | Accuracy indicators |
|---|---|---|
| Mirror temperature module | Automatic cooling control | ±0.5℃ |
| Pressure measurement module | Explosion-proof design | Grade 1.0 accuracy |
| Sample processing unit | Pressure stabilization and filtration | Flow rate 1-3L/min |
Key points of the operation process
- Sampling requirements: The pipeline temperature must be at least 3°C higher than the estimated hydrocarbon dew point, and the low temperature environment must be heated and kept warm
- Flow control: Maintain a constant flow of 1-3L/min during formal measurement
- Repeatability verification: The difference between two consecutive measurements must be ≤2.0°C to be valid
Application case: A West-East Gas Transmission Station uses an AMA-HCP3000 dew point meter. The hydrocarbon dew point is measured to be -15.2°C at a pressure of 5.0MPa. The maximum deviation of three repeated measurements is 1.3°C, which meets the standard requirements.
Background of Standard Evolution
Compared with the visual inspection method of GB/T 27895-2011, this standard introduces automated detection technology, reducing the manual judgment error from ±3℃ to ±2℃, and increasing explosion-proof safety requirements to adapt to the actual working conditions of the station.
Implementation Suggestions
- Regularly use propane standard gas (-42.1℃@101.325kPa) for instrument calibration
- For high hydrocarbon dew point samples (>0℃), the purge flow rate needs to be increased to 5L/min
- Establish a mirror contamination assessment system, and clean the mirror when the repeatability is >1.5℃

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

SY/T 5429-1991 in English
Small diameter flow moisture tester
1992-01-15 -

SY 5747-1995 in English
Safety rules for offshore oil construction projects
1995-11-18 -

SY 5272-1991 in English
Conventional drilling safety technology
1991-07-19 -

SY/T 6094-1994 in English
Evaluation Procedure for Lubricants Used in Drilling Fluids
1995-01-18 -

SY/T 5925-1994 in English
Selection Method of Chemical Tracer for Oilfield Water Injection
1994-08-04 -

SY/T 4069-1993 in English
Petroleum Construction Quality Inspection and Evaluation Standards oilfield steel container production (with the provisions stated)
1993-09-09 -

SY/T 5439-2003 in English
Pneumatic cement-discharging tanker
2003-03-18