WS 310.2-2016 in English
VALIDCentral sterile supply department(CSSD) Part 2:standard for operating procedure of cleaning,disinfection and sterilization
- Issued on:2016-12-27
- Implemented on:2017-06-01
- File Format:PDF
- Delivery:Via email within 1~3 business days
$243.00
WS 310的本部分规定了医院消毒供应中心的诊疗器械、器具和物品处理的基本要求、操作流程。
本部分适用于医院和为医院提供消毒灭菌服务的消毒服务机构。
Introduction
Analysis of the core content of the standard
This standard replaces WS 310.2-2009. The main technical changes include: adding processing requirements for precision instruments and luminal instruments, adjusting moist heat disinfection parameters, and refining the management of foreign medical devices, etc. 12 important revisions.
Comparison of key operating procedures
| Process links | 2009 version requirements | 2016 version updates |
|---|---|---|
| Moist heat disinfection | Temperature ≥ 90℃ Time ≥ 1min | Add A0 value ≥ 3000 indicator |
| Pressure steam sterilization | Pressure range is not clear | Precise range of 184.4-229.3kPa is specified |
| Foreign medical devices | Not separately specified | Add 4.7 special clauses |
Special Instrument Handling Specifications
Key Points for Luminal Instrument Operation
For lumen instruments with an inner diameter ≥ 2mm, the standard requires:
- A special cleaning brush must be used for lumen treatment (Appendix B.1)
- A pressure air gun is used to remove residual water during the drying phase (5.5.3)
- Downward exhaust pressure steam sterilization is prohibited (5.8.1.5)
Detailed Explanation of Sterilization Technical Parameters
Key Parameters for Pressure Steam Sterilization:
| Sterilizer Type | Temperature | Minimum time | Pressure range |
|---|---|---|---|
| Downward exhaust type | 121℃ | 20-30min | 102.8-122.9kPa |
| Pre-vacuum type | 132-134℃ | 4min | 201.7-229.3kPa |
Implementation recommendations
- Establish a special processing procedure for precision instruments and equip them with protective devices
- Regularly verify the A0 value calculation function of the washer-disinfector
- Two-person inspection system is implemented for external medical devices
- Hard containers are tested for sealing every month

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

WS 843-2025 in English
Medical equipment configuration standards for township health centers
2025-03-29 -

WS/T 845-2024 in English
Standards for bed transfer between the elderly care area and the medical area in medical and nursing institutions
2024-07-24 -

WS/T 594-2018 in English
Test methods of fly resistance to insecticides-A molecular method for detecting insensitive ace allele in the Musca domestica
2018-03-01 -

WS/T 508-2016 in English
Regulation for washing and disinfection technique of medical textiles in healthc are facilities