VALIDMethod for chemical analysis of titanium sponge, titanium and titanium alloys-Determination of multielement content-Inductively coupled plasma atomic emission spectrometry
- Issued on:2018-10-22
- Implemented on:2019-04-01
- File Format:PDF
- Delivery:Via email within 1~3 business days
$185.00
本标准规定了海绵钛、钛及钛合金中铝、硼、铋、钴、铬、铜、铁、铪、镁、锰、钼、铌、镍、铅、钯、钌、硅、锡、钽、钒、钨、钇、锌、锆含量的测定方法。
本标准适用于海绵钛、钛及钛合金中铝、硼、铋、钴、铬、铜、铁、铪、镁、锰、钼、铌、镍、铅、钯、钌、硅、锡、钽、钒、钨、钇、锌、锆含量的测定。
Introduction
Technical background of the standard
This standard was proposed by the National Technical Committee for Standardization of Nonferrous Metals (SAC/TC 243) and jointly formulated by 10 units including Baoti Group, filling the gap in the standard for simultaneous detection of multiple elements in titanium materials. The inductively coupled plasma emission spectrometer is used to achieve efficient determination of 23 impurity elements, which is 5-8 times more efficient than traditional chemical analysis.
Principle of the method
After the sample is dissolved in the hydrochloric acid-hydrofluoric acid-nitric acid system, it is excited in the ICP-AES plasma light source, and the content is calculated by measuring the intensity of the characteristic spectral lines of each element using the working curve method. Key parameters:
| Parameters | Requirements | Remarks |
|---|---|---|
| Plasma power | 1.0-1.5kW | Adjust according to element characteristics |
| Atomizing gas flow rate | 0.6-1.0L/min | Signal-to-noise ratio needs to be optimized |
| Observation height | 10-15mm | Axial observation |
Key operating points
1. Sample pretreatment
Use PTFE beakerDissolve the sample, strictly control the temperature: ≤70℃ for B and Si, other elements can be heated appropriately. Complete dissolution is marked by clear solution without suspended matter.
2. Standard solution configuration
Standard storage solution must meet the following requirements:
- Matrix matching: Use 99.99% high purity titanium as base
- Acidity control: HCl-HF-HNO₃ system consistent with the sample solution
- Concentration gradient: Configure the calibration points recommended in Table 3
Technology evolution comparison
| Dimension | Traditional method | This standard method |
|---|---|---|
| Detection efficiency | Single element determination in sequence | Simultaneous determination of 23 elements |
| Detection limit | 10⁻⁴% level | B/Y reaches 10⁻⁵% |
| Precision | RSD≈5% | RSD≤2% |
Implementation suggestions
Key points of quality control
- Each batch of tests must be accompanied by matrix blanks and certified standards
- Regularly check the nebulizer efficiency (≥98%)
- Spectral line interference correction: use interference coefficient method or MSF correction
Abnormal handling
When When the results of Zr and Hf are abnormal, it is recommended to:
- Check the amount of hydrofluoric acid added (2±0.1mL)
- Verify the sealing of the hydrofluoric acid resistant injection system

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

YS 41-1992 in English
Beryllium piece
1992-03-19 -

YS/T 63.6-2019 in English
Test method for carbonaceous materials used in the production of aluminium - Part 6: Determination of open porosity
2019-08-02 -

YS 102-1992 in English
Lead and zinc smelting enterprise energy consumption
1992-03-17 -

YS/T 10-1991 in English
Multifunctional unit for anode baking furnace
1991-03-22 -

YS/T 1050.1-2015 in English
Methods for chemical analysis of lead antimony concentrates—Part 1:Determination of lead content— Na2EDTA titration method
2015-04-30 -

YS/T 5231-2024 in English
Standard for geotechnical investigation of red mud yard
- -

YS/T 438-2001 in English
Determination method of physical properties of sandy alumina
2001-02-12 -

YS/T 64-1993 in English
Method for determination of resistivity of carbon products for aluminum electrolysis
1993-03-17 -

YS/T 421-2000 in English
PS plate aluminum substrate for printing
2000-03-29 -

YS/T 4-1991 in English
Ground form pneumatic sheller
1991-03-22