《GB/T 17674-2021原油中氮含量的测定 舟进样化学发光法》由TC355(全国石油天然气标准化技术委员会)归口,TC355SC10(全国石油天然气标准化技术委员会油气计量及分析方法分会)执行,主管部门为国家标准化管理委员会。
Introduction
Core changes in the standard revision
| Technical elements | 2012 edition | 2021 edition | Significance of the revision |
|---|---|---|---|
| Unit specification | Mixed use mass fraction/concentration | Unified as mg/kg | Improve data comparability |
| Preparation of standard solution | Volume method only | Add mass method | Improve accuracy |
| Calculation formula | 4 formulas | 8 formulas | Covering more application scenarios |
Key instrument configuration requirements
Boat injection system must meet the following requirements:
- Sample boat material: platinum or quartz, lined with quartz wool
- Injection speed controllable range: 150±10mm/min
- Sealing requirements: able to withstand high temperature of 1100℃
Combustion tube configuration differences:
- Single-stage furnace temperature 1100℃±25℃
- Dual-stage needs to have carrier gas/oxygen switching function
Analysis of key operation points
Establishing a standard curve
Note when using a 5μL syringe for injection:
- The volume method requires three rinses to eliminate bubbles
- The weighing accuracy of the mass method must reach 0.01mg
Typical problem solving
| Phenomenon | Cause | Solution |
|---|---|---|
| Residual coke in the sample boat | Insufficient residence time | Extend the time in the high temperature zone by ≥2min |
| Carbon deposition at the end of the combustion tube | Injection speed is too fast | Adjust to 150±10mm/min |
Quality Control Implementation Plan
QC sample selection should meet the following requirements:
- Nitrogen content covers the range of 40-10000mg/kg
- National primary standard materials should be used preferentially
- The analysis interval for each batch shall not exceed 20 samples
Precision control requirements:
Repeatability: r ≤ 0.087X (X is the mean)
Reproducibility: R ≤ 0.266X (X is the mean)

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

GB/T 8929-2006 in English
Crude petroleum - Ddtermination of water - Distillation method
2006-01-18 -

GB/T 25104-2019 in English
Automatic measurement of water content in crude oil—RF method
2019-08-30 -

GB/T 6531-1986 in English
Crude petroleum and fuel oils; Determination of sediment; Extraction method
1986-06-25 -

GB/T 11146-2009 in English
Crude petroleum - Determination of water - Coulometric Karl Fischer titration method
2009-03-16 -

GB/T 17606-2009 in English
Determination of sulfur in crude-oil by energy-dispersive X-ray fluorescence spectrometry
2009-04-08 -

GB/T 17280-2017 in English
Standard test method for distillation of crude petroleum―15-theoretical plate column
2017-12-29