WS/T 659-2019 in English
VALIDSafety management for multi-parameter patient monitor
- Issued on:2019-10-18
- Implemented on:2020-05-01
- File Format:PDF
- Delivery:Via email within 1~3 business days
$185.00
本标准规定了多参数监护仪(以下简称监护仪)临床使用期间安全管理的基本要求、检测的项目和方法。
本标准适用于医疗机构临床使用的监护仪的安全管理。
Introduction
Interpretation of the core content of the standard
As an important technical specification in the field of medical monitoring equipment, this standard has established a technical framework for the full life cycle management of multi-parameter monitors. Its outstanding features are reflected in:
| Management dimension | New standard requirements | Comparison with traditional practices |
|---|---|---|
| Detection cycle | ≤1 time/yearMandatory detection | No unified regulations |
| Leakage current control | 4 types of current classification control | Only focus on ground leakage current |
| Performance error | Blood oxygen±3%, blood pressure±4mmHg | Depends on manufacturer standards |
Detailed explanation of key technical indicators
Electrical safety testing system
The standard innovatively establishes a four-level current detection matrix:
- Ground leakage current: BF/CF type equipment is required to be ≤1mA (fault state)
- Patient auxiliary current: CF type equipment AC mode is strictly controlled to 0.01mA
- The detection method covers 6 fault simulation states
Performance parameter calibration specification
Take non-invasive blood pressure detection as an example:
- Static pressure detection range 0-260mmHg
- First detection error ≤3mmHg
- Repeatability detection requires 5 measurements
- Airtightness requirement ≤6mmHg/min
Implementation suggestions for medical institutions
Management structure construction
The standard requires the establishment of a three-level management organization:
- Decision-making level: hospital leaders bear leadership responsibilities
- Execution level: medical device management departments lead testing
- Operation level: using departments implement daily maintenance
Testing capability construction
Professional testing equipment is required:
- ECG simulators need to support arrhythmia signals
- Blood oxygen simulators should contain multiple oxygen dissociation curves
- Ambient temperature is controlled within the range of 5-40℃
Standard evolution analysis
This standard integrates the requirements of GB 9706.1 and GB 9706.25 for the first time, organically combining the general standards for medical electrical equipment with the special standards for ECG monitoring. Its technological breakthroughs are reflected in:
- New respiratory rate detection specifications (15-80 times/min)
- Detailed neonatal blood pressure detection requirements
- Establishment of a mandatory post-maintenance detection system

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

WS 843-2025 in English
Medical equipment configuration standards for township health centers
2025-03-29 -

WS/T 845-2024 in English
Standards for bed transfer between the elderly care area and the medical area in medical and nursing institutions
2024-07-24 -

WS/T 528-2016 in English
Technical specifications for small centralized water supply disinfection
2016-12-26 -

WS/T 594-2018 in English
Test methods of fly resistance to insecticides-A molecular method for detecting insensitive ace allele in the Musca domestica
2018-03-01