SN/T 5444-2022 in English
VALIDDetermination of prochloraz and its metabolites inplant-origin foodstuffs for export - LC-MS/MS
- Issued on:2022-03-14
- Implemented on:2022-10-01
- File Format:PDF
- Delivery:Via email within 5 business days
$330.00
Introduction
Technical Background and Standard Value
Prochloraz is widely used as a broad-spectrum fungicide in agricultural production, and its metabolites BTS44595, BTS44596 and BTS45186 also have biological activity. This standard establishes a detection method with a sensitivity of 0.01mg/kg in response to the increasingly stringent pesticide residue limit requirements in international trade, providing technical support for ensuring the compliance of exported food.
Core Innovation of the Method
| Technical Aspects | Traditional Methods | Improvements of This Standard |
|---|---|---|
| Extraction Method | Single Solvent Extraction | Acidified Acetonitrile Combined with QuEChERS/SPE Dual Cleanup System |
| Detection Sensitivity | 0.05mg/kg | 0.01mg/kg (Increased 5 Times) |
| Matrix Coverage | 3-5 Types of Food | 9 Types of Plant-derived Food (including special matrices such as tea and dendrobium) |
Analysis of key operation points
1. Differentiated pre-treatment scheme
Customized treatment is used for different food matrices: tea needs to add a rotary evaporation concentration step; high-fat samples (such as nuts) use GCB to remove pigment interference.
2. Optimization of mass spectrometry conditions
Use positive and negative ion switching mode to simultaneously detect: prochloraz and its metabolites (ESI+), 2,4,6-trichlorophenol (ESI-), and the dwell time is set to 40ms and 100ms respectively to ensure sensitivity.
Implementation recommendations
- Quality control: Each batch of samples should be subjected to a blank test and spike recovery test (recommended addition level 0.01-2.0 mg/kg)
- Equipment calibration: The ion abundance ratio should be verified with a matrix-matched standard solution before daily analysis (refer to Table 3 for deviation requirements)
- Data report: When the result is less than 0.1 mg/kg, retain four decimal places; when it is greater than or equal to 0.1 mg/kg, retain three significant figures
Technology evolution analysis
Compared with the SN/T 2325-2009 version of the method, the progress of this standard is reflected in:
① The introduction of graphitized carbon black-amino composite column purification technology increases the recovery rate to 75-109%;
② The total amount of metabolites can be converted through the conversion coefficient (1.157-1.908), which is more in line with international regulatory trends.

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

SN 0283-1994 in English
\"Method for Inspection of Dichlorodipyridinol Residues in Exported Poultry Meat\"
1994-02-18 -

SN/T 0244-1994 in English
\"General Rules for Inspection and Certification of Imported Mechanical and Electrical Instruments\"
1993-08-01 -

SN 0346-1995 in English
Method for the determination of α-naphthylacetic acid residues in vegetables for export
- -

SN/T 3576-2013 in English
Detection of genetically modified components-Method for soybean using PCR-DHPLC
2013-03-01 -

SN/T 0736.2-2011 in English
Test Method of Import and Export Fertilizers. Part 2: Determination of Moisture
2011-09-09