《GB/T 22062-2020显微镜 目镜分划板》由TC103(全国光学和光子学标准化技术委员会)归口,TC103SC2(全国光学和光子学标准化技术委员会显微镜分会)执行,主管部门为中国机械工业联合会。
Introduction
Background and significance of standard revision
GB/T 22062-2020 is the latest national standard for microscope eyepiece graticules, replacing the 2008 version of the standard. The technical content refers to ISO 9344:2016 (non-equivalent adoption). This standard specifies the manufacturing requirements for cross-reticles used for optical measurement, which directly affects the measurement accuracy of microscopes.
Comparison of major technical changes
| Changed items | GB/T 22062-2008 | GB/T 22062-2020 |
|---|---|---|
| Type I graticule size | Not separately specified | New Φ19/Φ21 diameter series |
| Reference standard | Footnote f in Table 2 | Delete footnote and integrate it into the main text |
| Fringe detection | Area calculation is not specified | Added a new diameter-fringe level correspondence table |
Detailed explanation of core technical requirements
3.1 Graticule size specification
The standard specifies the size requirements for two types of graticules:
- Type I graticule: two new diameters of Φ19-0.1 and Φ21-0.1, with thickness maintained at 1.0±0.1mm
- Type II graticule: maintain diameters of Φ19-0.033/Φ21-0.033/Φ26-0.033, with chamfers of 1.5±0.2mm
3.2 Material defect detection
According to GB/T11162 and GB/T1185-2006, the key indicators include:
Case: For the Φ21mm graticule, the bubble defect requirement is that area 1 allows 1.5×√0.016mm² defects, which is equivalent to allowing 2 bubbles with a diameter of ≤0.14mm.
Implementation Suggestions
- Selection Guide: For high-precision measurement, type II graticules are preferred, as they have a tighter diameter tolerance (0.033mm vs 0.1mm)
- Quality Control: It is recommended to use an interferometer to detect surface shape errors to ensure ≤10' parallelism
- Label Management: The graticule body or packaging must be marked with "Compliant with GB/T22062-2020"

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

GB/T 44276.1-2024 in English
Microscopes—Cover glasses—Part 1:Dimensional tolerances, thickness and optical properties
2024-08-23 -

GB/T 42886-2023 in English
Microscopes—Microscopes with digital imaging displays—Information provided to the user regarding imaging performance
2023-08-06 -

GB/T 2985-2008 in English
Biological microscope
2008-07-28 -

GB/T 36529-2018 in English
Stereomicroscopes—Information provided to the user
2018-07-13 -

GB/T 24665-2009 in English
Polarizing microscope
2009-11-15 -

GB/T 9246-2008 in English
Microscopes—Oculars (eyepieces)
2008-06-20