Analysis of standard technical framework
| Technical elements | Gas chromatography (GC) | Gas chromatography-mass spectrometry (GC-MS) |
|---|---|---|
| Main functions | Quantitative analysis | Qualitative confirmation |
| Detector type | Flame ionization detector (FID) | Electron impact source (EI) |
| Characteristic indicators | Chromatographic peak area | Retention time + characteristic ion fragments (m/z 429*,414,324) |
Key Operation Points
Sample pretreatment uses a combination of vortex oscillator and centrifuge technology:
- Solid samples need to be ground to 100 mesh or above
- Methanol extraction time is strictly controlled at 10±0.5 minutes
- Centrifugation parameters are not less than 4000r/min×5min
Mass spectrometry qualitative judgment criteria
Taking the mass spectrum in Appendix B as an example, a positive judgment must meet the following requirements at the same time:
- Base peak m/z 429 relative intensity deviation ≤±10%
- Secondary characteristic peak m/z 414Relative intensity deviation ≤±15%
- Retention time window±1% (about±0.2min)
Implementation suggestions
The laboratory should establish a dual system verification mechanism:
| Quality control link | Control requirements |
| Standard solution | Each batch of inspection must be accompanied by Qualitative quality control standard working solution (0.01mg/mL) |
| Instrument calibration | DB-5MS chromatographic column needs to be regularly column efficiency tested |

Loading PDF document...
Error loading PDF. Please make sure the file is valid and try again.
We also recommend
-

GB/T 31868-2015 in English
Technical requirements for anti-counterfeiting ofD9ing matrix code generator
2015-09-11 -

GB/T 23865-2009 in English
The scales photography method
2009-03-09 -

GB/T 29347-2012 in English
Examination standard for cartridge case marks in firearms discipline of forensic science
2012-12-31 -

GB/T 34919-2017 in English
Code of practice for examination of printed documents in forensics
2017-12-29 -

GB/T 37226-2018 in English
Criterion of forensic science human fluorescence STR multiplex amplification reagent
2018-12-28 -

GB/T 40991-2021 in English
Collecting and packaging methods for trace evidence
2021-11-26 -

GB/T 39874-2021 in English
Examination methods for bromazepam in suspected drugs—Gas chromatography and gas chromatography-mass spectrometry
2021-04-30 -

GB/T 29635-2013 in English
GC and GC-MS examination methods for heroin in suspected drug
2013-07-19